Recall Z-0038-2013— Medtronic Sofamor Danek USA Inc
Class II · Terminated
Class II FDA medical device recall by Medtronic Sofamor Danek USA Inc, initiated 2012-08-30, terminated 2013-08-15. It names 3 FDA 510(k) clearances: K081038, K063100, K021461. Product: PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.. Reason: Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length..
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-08-30
- Terminated
- 2013-08-15
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
3 on FDA's record- K081038 — ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM
- K063100 — ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM, TRANSLATIONAL PLATE
- K021461 — ATLANTIS VISION ANTERIOR CERVICAL PLATE SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.
Reason for recall
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
Data sourced from openFDA. This site is unofficial and independent of the FDA.