Recall Z-0037-2022— CardioQuip, LLC

Class II · Ongoing

Class II FDA medical device recall by CardioQuip, LLC, initiated 2021-05-12. It names one FDA 510(k) clearance: K102147. Product: CardioQuip, Modular Cooler-Heater 1000(m),Compact Unit, Model:MCH-1000(m),CE 2797, UDI: (01)00860000846110. Reason: Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices..

Recalling firm
CardioQuip, LLC
Classification
Class II
Status
Ongoing
Initiated
2021-05-12
Firm location
College Station, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CardioQuip, Modular Cooler-Heater 1000(m),Compact Unit, Model:MCH-1000(m),CE 2797, UDI: (01)00860000846110

Reason for recall

Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.

Data sourced from openFDA. This site is unofficial and independent of the FDA.