Recall Z-0037-2019— Oridion Medical 1987 Ltd.

Class II · Terminated

Class II FDA medical device recall by Oridion Medical 1987 Ltd., initiated 2018-07-31, terminated 2022-09-20. FDA's recall record names no specific 510(k) clearance for this event. Product: CAPNOSTREAM 20P (INTL) N, Part Number CS08652-01. Reason: The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off..

Recalling firm
Oridion Medical 1987 Ltd.
Classification
Class II
Status
Terminated
Initiated
2018-07-31
Terminated
2022-09-20
Firm location
Jerusalem, Israel

Product description

CAPNOSTREAM 20P (INTL) N, Part Number CS08652-01

Reason for recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

Data sourced from openFDA. This site is unofficial and independent of the FDA.