Recall Z-0035-2025— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2023-04-19. It names 2 FDA 510(k) clearances: K153702, K113125. Product: Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1. Reason: During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails..

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2023-04-19
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Reason for recall

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.