Recall Z-0035-2018— St Jude Medical Inc.

Class II · Terminated

Class II FDA medical device recall by St Jude Medical Inc., initiated 2017-08-28, terminated 2020-07-08. FDA's recall record names no specific 510(k) clearance for this event. Product: PM1226 ACCENT ST MRI SR RF and PM2222 ACCENT ST These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications..

Recalling firm
St Jude Medical Inc.
Classification
Class II
Status
Terminated
Initiated
2017-08-28
Terminated
2020-07-08
Firm location
Sylmar, CA, United States

Product description

PM1226 ACCENT ST MRI SR RF and PM2222 ACCENT ST These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.

Reason for recall

New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.

Data sourced from openFDA. This site is unofficial and independent of the FDA.