Recall Z-0033-2026— Medtronic MiniMed, Inc.
Class II · Ongoing
Class II FDA medical device recall by Medtronic MiniMed, Inc., initiated 2025-05-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic, Simplera Sensor, REF: MMT-5100JD1. Reason: The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection..
- Recalling firm
- Medtronic MiniMed, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-05-07
- Firm location
- Northridge, CA, United States
Product description
Medtronic, Simplera Sensor, REF: MMT-5100JD1
Reason for recall
The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.