Recall Z-0032-2026— Beckman Coulter Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2025-08-04. It names one FDA 510(k) clearance: K242190. Product: DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer. Reason: On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-08-04
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reason for recall

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.