Recall Z-0027-2024— Bard Peripheral Vascular Inc
Class II · Ongoing
Class II FDA medical device recall by Bard Peripheral Vascular Inc, initiated 2023-08-10. It names one FDA 510(k) clearance: K133948. Product: Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016. Reason: Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding..
- Recalling firm
- Bard Peripheral Vascular Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-08-10
- Firm location
- Tempe, AZ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016
Reason for recall
Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.