Recall Z-0026-2014— Boston Scientific CRM Corp

Class II · Terminated

Class II FDA medical device recall by Boston Scientific CRM Corp, initiated 2013-08-29, terminated 2014-04-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support.. Reason: The performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to December 2009, may be compromised over time, causing increased current drain that can lead to premature battery depletion..

Recalling firm
Boston Scientific CRM Corp
Classification
Class II
Status
Terminated
Initiated
2013-08-29
Terminated
2014-04-10
Firm location
Saint Paul, MN, United States

Product description

Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support.

Reason for recall

The performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to December 2009, may be compromised over time, causing increased current drain that can lead to premature battery depletion.

Data sourced from openFDA. This site is unofficial and independent of the FDA.