Recall Z-0025-2025— Beckman Coulter, Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter, Inc., initiated 2024-04-18. It names one FDA 510(k) clearance: K221225. Product: Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.. Reason: When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians..

Recalling firm
Beckman Coulter, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-04-18
Firm location
Chaska, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reason for recall

When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.