Recall Z-0024-2018— St. Jude Medical, Inc.
Class II · Terminated
Class II FDA medical device recall by St. Jude Medical, Inc., initiated 2017-09-12, terminated 2020-04-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Proclaim DRG Implantable Pulse Generator Model 3664. Reason: Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations..
- Recalling firm
- St. Jude Medical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-09-12
- Terminated
- 2020-04-16
- Firm location
- Plano, TX, United States
Product description
Proclaim DRG Implantable Pulse Generator Model 3664
Reason for recall
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.