Recall Z-0020-2018— Philips Electronics North America Corporation

Class II · Terminated

Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2017-01-18, terminated 2020-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Philips Healthcare Brilliance iCT SP X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.. Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient..

Recalling firm
Philips Electronics North America Corporation
Classification
Class II
Status
Terminated
Initiated
2017-01-18
Terminated
2020-02-26
Firm location
Andover, MA, United States

Product description

Philips Healthcare Brilliance iCT SP X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason for recall

During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.