Recall Z-0019-2025— Outset Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Outset Medical, Inc., initiated 2024-08-01. It names one FDA 510(k) clearance: K233335. Product: Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.. Reason: Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath..
- Recalling firm
- Outset Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-08-01
- Firm location
- San Jose, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.
Reason for recall
Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.