Recall Z-0017-2018— Philips Electronics North America Corporation
Class II · Terminated
Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2017-01-18, terminated 2020-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Philips Healthcare Ingenuity CT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.. Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient..
- Recalling firm
- Philips Electronics North America Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-01-18
- Terminated
- 2020-02-26
- Firm location
- Andover, MA, United States
Product description
Philips Healthcare Ingenuity CT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for recall
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
Data sourced from openFDA. This site is unofficial and independent of the FDA.