Recall Z-0016-2026— Getinge Disinfection Ab

Class II · Ongoing

Class II FDA medical device recall by Getinge Disinfection Ab, initiated 2025-08-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Getinge 88-Series Washer- Disinfector. Model Number: 88-5.. Reason: Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered..

Recalling firm
Getinge Disinfection Ab
Classification
Class II
Status
Ongoing
Initiated
2025-08-29
Firm location
Vaxjo, N/A, Sweden

Product description

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Reason for recall

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Data sourced from openFDA. This site is unofficial and independent of the FDA.