Recall Z-0014-2018— Handicare AB

Class II · Terminated

Class II FDA medical device recall by Handicare AB, initiated 2016-07-28, terminated 2018-12-17. FDA's recall record names no specific 510(k) clearance for this event. Product: Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009. Reason: A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye..

Recalling firm
Handicare AB
Classification
Class II
Status
Terminated
Initiated
2016-07-28
Terminated
2018-12-17
Firm location
Lulea, Sweden

Product description

Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

Reason for recall

A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.