Recall Z-0009-2026— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-09-03. It names 2 FDA 510(k) clearances: K130842, K041949. Product: Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;. Reason: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2025-09-03
Firm location
Best, N/A, Netherlands

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Reason for recall

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Data sourced from openFDA. This site is unofficial and independent of the FDA.