Recall Z-0008-2014— Draeger Medical, Inc.

Class I · Terminated

Class I FDA medical device recall by Draeger Medical, Inc., initiated 2013-08-14, terminated 2014-09-23. It names one FDA 510(k) clearance: K042419. Product: Draeger Fabius GS Premium, Fabius OS, Fabius Tiro, Fabius Tiro D-M Product Usage: used to administer anesthesia and ventilation to patients during surgical procedures.. Reason: During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit housing was not maintained for some from a specific batch..

Recalling firm
Draeger Medical, Inc.
Classification
Class I
Status
Terminated
Initiated
2013-08-14
Terminated
2014-09-23
Firm location
Telford, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Draeger Fabius GS Premium, Fabius OS, Fabius Tiro, Fabius Tiro D-M Product Usage: used to administer anesthesia and ventilation to patients during surgical procedures.

Reason for recall

During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit housing was not maintained for some from a specific batch.