Recall Z-0002-2026— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Class II · Ongoing
Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-09-03. It names 2 FDA 510(k) clearances: K130842, K041949. Product: Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;. Reason: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted..
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-09-03
- Firm location
- Best, N/A, Netherlands
510(k) clearances named in this recall
2 on FDA's record- K130842 — ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
- K041949 — ALLURA XPER FD10
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Reason for recall
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.